TDLAS Gas Detection Modules for Industrial Integration

TDLAS Modules
Gas Analysis Integration for Environmental and Industrial Monitoring

The Sartec Photoelectronics H2S module uses tunable diode laser absorption spectroscopy (TDLAS) for hydrogen sulfide measurement in conditioned, flow-through gas samples. It provides an RS485 communication interface using Modbus for integration into gas analysis instruments.

TDLAS Modules for detecting gases at extremely low concentrations with application-specific sensitivity, used in environmental monitoring, petrochemical, and healthcare industries

Application review

This page describes the H2S flow-through TDLAS module. Review the required H2S concentration range, background gases, sample conditioning and instrument requirements for the project. Other gases require separate optical configuration, calibration and model-specific documentation.

TDLAS (Tunable Diode Laser Absorption Spectroscopy) Gas Detection Module in an industrial environment for trace gas detection and monitoring.

Technical Specifications

Target Gas Hydrogen Sulfide (H2S)
Gas Supply Flow
Principle of Measurement TDLAS (Tunable Diode Laser Absorption Spectroscopy)
Measuring Range 0-200 ppm (or F.S. ≥200 ppm available)
Linearity ≤±1% F.S.
Repeatability ≤±1% F.S.
Span Drift ≤±2% F.S. / 6 months
Detection Limit 2 ppm
Response Time (T90) ≤30s
Communication Connection RS485
Communication Protocol Modbus
Suggested Gas Flow 1-5 L/min
Sample Gas Dust, moisture, and oil-free gases
Housing Material Metal plate
Gas Pipe Connector 4/6mm (PTFE)
Operating Voltage 12V ± 1V
Dimensions 200mm (L) x 175mm (W) x 76mm (H)

Published H2S configuration

The specification table on this page is for the H2S configuration documented in H2S TDLAS Module Instructions V3.0, page 5. It lists 12 V ±1 V, RS485 using Modbus, 0–200 ppm and T90 ≤30 s. Other full-scale ranges of 200 ppm or above are described as available on request. Confirm the selected configuration before ordering.

View H2S TDLAS Module Instructions V3.0

OEM integration requirements

Review the sampling arrangement, inlet and outlet identification, gas compatibility and required sample conditioning before installation. The manual specifies conditioned gas and a stable installation without obvious vibration. Housing material alone does not establish an outdoor ingress-protection rating or suitability for an explosive atmosphere.

Confirm the wiring, power-supply capacity and communication settings against the applicable manual for the delivered module. Define alarm handling, calibration, maintenance and final instrument acceptance tests as part of the system design.

RFQ Information for Gas Module Review

For gas module selection, provide the target gas, measurement range, optical path or sampling method, response-time requirement, output interface, installation environment, and calibration expectations. Sartec Photoelectronics can review the application requirements before recommending a suitable module configuration.

Technical Support

We provide comprehensive technical support to ensure that your gas detection module matches the intended operating conditions. Our services include:
•Installation guidance and setup assistance.
•Troubleshooting for defined performance requirements.
•Ongoing support for maintenance, upgrades, and calibration.

Product Line Link

FAQ – TDLAS H2S Modules

What gases can TDLAS H2S Modules detect?

This specification applies to hydrogen sulfide (H2S). Other target gases require their own optical configuration, calibration and documentation.

What is the detection range of the TDLAS H2S Modules?

The manual publishes a 0–200 ppm example and describes other full-scale ranges of 200 ppm or above as available. Confirm the selected range and performance in the quotation and datasheet.

What is the response time of TDLAS H2S Modules?

The published H2S configuration lists T90 ≤30 s. Confirm the sample conditions and final sampling-system response.

Can these modules be integrated with other monitoring systems?

The published H2S configuration provides RS485 communication using Modbus. Verify settings and wiring against the applicable manual.

What are the power requirements for the TDLAS H2S Modules?

The manual lists 12 V ±1 V. Confirm power-supply capacity and wiring for the delivered configuration before connection.

Are the TDLAS H2S Modules suitable for outdoor use?

Confirm enclosure, environmental limits and required approvals for the selected module and final instrument. This page does not establish an IP rating, hazardous-area approval or certification of the complete safety function.

How often do the modules need to be calibrated?

Define calibration and maintenance for the application using the applicable manual and operating conditions. The published span-drift figure is within ±2% of full scale over six months; it does not by itself establish a calibration interval.

 

Get a Quote / Request Model Confirmation

To receive an accurate quote or to request model confirmation for your specific application, please provide us with the following information:

  • Measurement Range: Please specify the required measurement range for your gas detection module. This helps us determine a suitable module based on your concentration needs.
  • Background Gases: Indicate any background gases that might be present in your environment. This ensures that the module is calibrated to account for interference from other gases.
  • Accuracy Requirements: Let us know the level of accuracy needed for your application. Different modules offer varying degrees of precision based on your measurement requirements.
  • Environmental Conditions: Provide details about the environment where the module will be used (e.g., temperature range, humidity, pressure, indoor/outdoor settings). This will help us recommend a module that performs optimally under those conditions.

Once we receive these details, our engineering team will review your requirements and recommend the most suitable model, ensuring it meets your detection and operational needs.

How to Get Started:

Simply fill out the contact form below or email us directly with the information listed above. Our team will get back to you with a tailored quote or product recommendation.

TDLAS gas analysis module engineering review notes

TDLAS module evaluation should focus on the target absorption line, optical path, pressure and temperature conditions, response time, and the interface required for process or instrument integration.

What information should be prepared before RFQ?

  • Target gas and absorption line selection
  • Open-path, extractive or cell-based optical configuration
  • Pressure, temperature and background gas conditions
  • Response time, output interface and calibration expectations

Can the module be selected by gas name alone?

No. The same gas can require different optical path, range, response time, sampling and output-interface choices. Sartec Photoelectronics reviews the measurement conditions before suggesting a suitable configuration.

What does the integration review focus on?

The review focuses on measurement method, operating environment, calibration plan, output interface, mechanical constraints and project documentation requirements.